Chapter 03 · Foundations

Parameters, units & reference ranges

Know what was measured—and what was assumed.

16 min read5 questions · 2 synthetic casesIndependent clinical review pending

By the end of this chapter

  • Read conventional units correctly
  • Distinguish saturation from partial pressure
  • Apply specimen-specific and laboratory-specific limits

Visual physiology

Three oxygen observations—not synonyms

  1. PaO₂ · mmHg

    Arterial plasma oxygen pressure

  2. SaO₂ · %

    Arterial haemoglobin saturation; method matters

  3. SpO₂ · %

    Pulse-oximeter estimate of saturation

Tends to lower pHTends to raise pH↑ CO₂ · ↓ HCO₃⁻↓ CO₂ · ↑ HCO₃⁻pH reflects the net effect of all processes
Pressure, arterial saturation and pulse estimates answer different questions. Report-specific intervals and measured versus calculated methods must remain visible.Original educational schematic · Independent clinical and visual review pending.

Reference intervals are context, not verdicts

Typical adult arterial teaching intervals are pH 7.35–7.45, PaCO₂ 35–45 mmHg and bicarbonate about 22–26 mmol/L. PaO₂ is often taught around 80–100 mmHg in a young adult at sea level, but it changes with age, altitude and oxygen delivery. The laboratory’s validated interval and clinical setting take precedence over a textbook range.

Electrolytes, lactate, hemoglobin and albumin add separate physiological information. Bicarbonate on the gas is usually calculated; chemistry total CO₂ is measured. Base excess estimates a nonrespiratory acid–base component under standardized conditions, but is not independent evidence when derived from the same gas. A value within physical plausibility bounds is not necessarily normal.

Units that change the answer

1 kPa ≈ 7.5006 mmHg · FiO₂ fraction = percentage ÷ 100

Pressure units must be standardized before applying a mmHg-based formula. A pCO₂ of 5.3 kPa is about 39.8 mmHg, not severe hypocapnia. For the P/F ratio use FiO₂ 0.40 rather than 40. Bicarbonate in mmol/L and mEq/L is numerically equal for this monovalent ion; that shortcut does not apply to every substance.

Albumin 20 g/L is 2.0 g/dL. An albumin-correction formula written for g/dL must not receive 20 without conversion. Be explicit about whether the anion-gap convention includes potassium. Compare an AG calculated without K⁺ only with a matching reference interval.

pO₂, SaO₂ and SpO₂ are different observations

PaO₂ is dissolved oxygen pressure in arterial plasma. SaO₂ describes arterial hemoglobin oxygen saturation; it may be measured with co-oximetry or calculated from a dissociation curve. SpO₂ is a pulse-oximeter estimate and must not be relabelled SaO₂. Dyshemoglobins and instrument assumptions can make calculated saturation misleading.

Use a reliably supplied report-specific saturation interval when available, with its method and source. Without one, any teaching interval must be labelled illustrative, not reported by the laboratory. Venous saturation reflects extraction and is not judged with an arterial interval. Capillary pO₂ is not PaO₂ and cannot safely drive arterial oxygenation calculations here.

Glucose, dyshaemoglobins and co-oximetry

Many blood-gas analyzers report whole-blood glucose. It is useful for rapid context, for example when an elevated-gap acidosis raises the question of ketoacids, but point-of-care whole-blood and laboratory plasma measurements can differ, and haematocrit, sample delay and analyzer method affect the value. A markedly abnormal glucose is a clinical finding in its own right: when clinically indicated, urgent assessment and treatment must not be delayed while awaiting laboratory confirmation. Confirm unexpected or decision-relevant results with the laboratory method when the situation allows, and interpret them alongside ketones and the clinical picture; glucose alone does not establish the cause of an acidosis.

Carboxyhaemoglobin (COHb) and methaemoglobin (MetHb) cannot carry oxygen normally, yet PaO₂ may be normal because it reflects dissolved oxygen rather than haemoglobin binding. A standard two-wavelength pulse oximeter is misled differently by each: with raised COHb it typically overestimates saturation, whereas methaemoglobinaemia drives the SpO₂ reading towards approximately 85%, overestimating or underestimating depending on the actual oxygen saturation. The useful comparison is therefore between SpO₂ and a measured arterial saturation, not between SpO₂ and PaO₂; an unexplained saturation discrepancy is a cue for co-oximetry. Co-oximetry, which measures haemoglobin species spectrophotometrically, is needed when carbon monoxide exposure, oxidising drugs or toxins, unexplained cyanosis, or a gap between SpO₂ and measured SaO₂ is suspected. Calculated SaO₂ assumes normal haemoglobin and is unreliable in these situations. Interpret COHb with exposure timing, oxygen already given and smoking history, using the laboratory’s own intervals.

Sampling and handling errors

Air bubbles equilibrate with the sample: pCO₂ falls, pH rises and pO₂ moves towards room-air values (about 150 mmHg), which can falsely raise a low PaO₂ or lower a high one. Expel bubbles promptly and cap the syringe. Delayed analysis allows continued cell metabolism, which tends to lower pO₂ and glucose and raise pCO₂ and lactate; leukocytosis or thrombocytosis can accelerate this. Follow local limits on time to analysis and whether to cool samples.

Excess liquid heparin dilutes the sample and can lower pCO₂ and bicarbonate; inadequate mixing, clots, haemolysis (which can raise potassium) and unintended venous sampling all distort results. An unexpected or internally inconsistent panel should prompt a check of specimen type, timing, handling and FiO₂ before interpretation, and repeat sampling when clinically appropriate. Arterial, venous and capillary samples remain distinct throughout.

Worked example

Normalize before calculating

  1. 1.Synthetic arterial pCO₂ = 5.3 kPa. Multiply by 7.5006 to obtain about 39.8 mmHg.
  2. 2.FiO₂ 40% becomes 0.40 for oxygenation calculations.
  3. 3.Albumin 20 g/L becomes 2.0 g/dL for the conventional gap correction.
  4. 4.Record the original printed units; conversion is a calculation, not permission to overwrite a source report.

Common mistakes & exam tips

  • Never insert oxygen percentage 40 into a formula expecting 0.40.
  • Do not apply arterial intervals to venous or capillary pO₂.
  • An exam range is an explicitly stated convention, not a universal laboratory standard.

Check your understanding

Five questions to consolidate the chapter

015.3 kPa is approximately how many mmHg?

02FiO₂ 35% is entered as which fraction?

03Which is an estimate from pulse oximetry?

04Albumin 30 g/L equals…

05Which interval should guide a reported laboratory result?

Reason through two clinical cases

SYNTHETIC CASEBeginner

A unit trap

Synthetic arterial gas printed using kPa, while an exam formula uses mmHg.

pH 7.40 · pCO₂ 5.3 kPa · HCO₃⁻ 24 mmol/L

SYNTHETIC CASEIntermediate

The wrong oxygen source

A synthetic capillary gas is submitted with an oxygen-delivery setting.

Capillary pO₂ 55 mmHg · FiO₂ 0.30

The essentials to take away

  • Units and provenance are part of every value.
  • Saturation and oxygen pressure are not interchangeable.
  • Laboratory intervals and specimen type govern comparison.